Evaluating Water Vapor Permeance Measurement Techniques for Highly Permeable Membranes

2015-12-13 06:46BuiDucThuanWongYonghuiChuaKianJonandNgKimChoon
Computers Materials&Continua 2015年8期
关键词:粪尿排泄物猪粪

Bui Duc Thuan,Wong Yonghui,Chua Kian Jonand Ng Kim Choon

Evaluating Water Vapor Permeance Measurement Techniques for Highly Permeable Membranes

Bui Duc Thuan1,2,Wong Yonghui2,Chua Kian Jon2and Ng Kim Choon2

The cup method and dynamic moisture permeation cell(DMPC)method are two common techniques used to determine the water vapor permeation properties of a membrane.Often,ignoring the resistance of boundary air layers to the transport of water vapor results in the water vapor permeance of the membrane being underestimated in practical tests.The measurement errors are higher with highly permeable membranes.In this study,the two methods were simulated using COMSOL Multiphysics platform and the extent of the error was evaluated.Initial results showed that the error is equally high in both methods.With the correction for the still air gap,the cup method produces a relatively reduced error.In the DMPC method,reducing the error caused by the boundary air layer by increasing the sweep speed can produce higher instrument error.Highly accurate and precise instrument is needed for DMPC method;however,its error is still higher than that in the cup method.Simulations also show that lowering the test pressure is favorable to both methods.

cup method simulation,dynamic moisture permeation cell method simulation,water vapor permeability

1 Introduction

The use of membranes for applications involving the removal or transfer of high amount of water vapor has driven the need for the evaluation of water vapor permeance of membranes with high water vapor permeance(upto6.8×10−6mol/m2.s.Pa[Xing,Rao,TeGrotenhuis,Canf i eld,Zheng,Winiarski and Liu(2013)])and high selectivity.Such applications include air dehumidif i cation[Yang,Yuan,Gao and Guo(2015);Metz,Van de Ven,Potreck,Mulder and Wessling(2005)],membrane heat and vapor recovery[Zhang and Jiang(1999)]and vapor/gas separation[Lin,Thompson,Serbanescu-Martin,Wijmans,Amo,Lokhandwala and Merkel(2012);Krull,Fritzmann and Melin(2008);Metz,Van De Ven,Mulder and Wessling(2005);Roy,Hussain and Mitra(2013);Scovazzo(2010);Sijbesma,Nymeijer,van Marwijk,Heijboer,Potreck and Wessling(2008)].

There are currently two widespread methods,the cup method[ASTM(2014)]and the dynamic moisture permeation cell(DMPC)method[ASTM(2009)],used for measuring water vapor transmission rates(WVTR)and thus water vapor permeance.Each method has its own variations[Metz,Van de Ven,Potreck,Mulder and Wessling(2005);Gennadios,Weller and Gooding(1994);Huang(2008);Huang and Qian(2008);Zhang(2006)],advantages and disadvantages[Huang and Qian(2008);McCullough,Kwon and Shim(2003)].

In the cup method,a membrane of a specif i c area covers an upright cup(Figs.1 a)or inverted cup(Fig.1 b),with the cup being filled with desiccant,water or salt solution to keep the relative humidity(RH)inside the cup fixed at a certain value.The cup is placed inside a temperature and humidity chamber with air or nitrogen atmosphere at controlled RH.In order to refresh the air/gas outside the cup,the air/gas in the chamber is circulated at a speed higher than 152 m/min as recommended in ASTM E96-14[ASTM(2014);McHugh,Avena-Bustillos and Krochta(1993)].In this cup method,the change in the mass of the cup’s content is monitored and used to quantify the water vapor permeation of the membrane.The advantage of this method is that it employs simple apparatus to conduct the test.

In the dynamic moisture permeation cell(DMPC)method,one side of the membrane is kept at fixed RH by blowing a fast-f l owing feed stream of air or gas over the membrane(Figs.1 c).On the other side of the membrane,a sweep air,helium or nitrogen is passed over.The fluid flows at the opposing sides of the membrane can be countercurrent or concurrent.They can be blown from one end to the opposite end of a rectangular membrane(Fig.1 c),or radially from the center of a circular membrane surface to the circumference or in the opposite direction[Metz,Van de Ven,Potreck,Mulder and Wessling(2005)].Water vapor diffuses through the membrane from a space with higher RH to one with lower RH,causing a change in RH of the sweep stream.The RH change is measured and used to calculate the permeability characteristic of the membrane.This method can be coupled with a gas chromatograph system to determine the permeability and selectivity of several gases or vapors at the same time,though the apparatus setup and operations are more complicated[Xing,Rao,TeGrotenhuis,Can field,Zheng,Winiarski and Liu(2013)].There are also setups,which are hybrids of the cup and DMPC methods,reported in literature[Huang(2008);Zhang(2006)].

Figure 1:Some examples of experimental setups to determine water vapor permeance:(a)upright cup method;(b)inverted cup method and;(c)counter-current f l ow DMPC method

For the cup method,the WVTR(mol/m2.s)is given as[ASTM(2014)]:

where Δmis the change in mass of the cup(g),tis the time taken for that change in mass(s),Ais the area of the membrane(m2),andMwis the molecular weight of water(g/mol).

For the DMPC method,it is determined from[ASTM(2009)]:

where δRHis the change in relative humidity between the incoming and outgoing stream of the sweep air/gas,Psis the saturation pressure of water vapour(Pa),Vis the volumetric f l ow rate(m3/s),R is the universal gas constant(J/mol.K),andTis the temperature of the measurement(K).

In order to quantify the ease with which water vapor can go through a membrane,vapor permeance(k,mol/m2.s.Pa)is employed and def i ned as the amount of water vapor that goes across a unit area of the membrane under a unit water vapor transmembrane pressure.With the assumption that the air resistance to the water vapor transport is negligible,the apparent water vapor transmembrane pressure,and thus apparent permeance,can be determined by

where ΔPw,appis the apparent water vapor transmembrane pressure(Pa),the driving force for water transmission through the membrane[ASTM(2014)].

For the cup method,the apparent water vapor transmembrane pressure is the water vapor pressure difference between the desiccant/water/salt solution surface inside the cup and the air/gas outside the cup.For the DMPC method,generally,the water vapor transmembrane pressure is the average water vapor pressure difference between feed stream and sweep stream[Metz,Van de Ven,Potreck,Mulder and Wessling(2005);ASTM(2009)].

For both methods,treating the resistance of air as negligible leads to an underestimation of water vapor permeance and causes a certain error for the measurement[Metz,Van de Ven,Potreck,Mulder and Wessling(2005);Gennadios,Weller and Gooding(1994);McHugh,Avena-Bustillos and Krochta(1993);Hu,Topolkaraev,Hiltner and Baer(2001)].The error is small for a low permeance membrane,whose resistance is much larger than the resistance of air on both sides of the membrane.The error is higher for a high permeance membrane if no correction is appropriately applied.So far,only the resistance of still air has been determined and considered for permeance calculations in the cup method[ASTM(2014)].The resistance of a moving gas and how it affects the measurement error are still not determined and evaluated in both methods.

Water vapor permeances obtained experimentally are not actual but apparent values.Therefore,although experimental analysis and comparison of the two methods have been done[Gennadios,Weller and Gooding(1994)],the extent of the measurement errors due to air resistance in both methods is not well-known.In particular,the error in the DMPC method,in which no applicable correction has been reported,is completely unknown.Therefore,what is the most appropriate method for a certain membrane and how to minimize the error in each method are questions yet to be addressed.

In this study,computational analysis was utilized to simulate the extent of the errors due to air resistance in a cup setup and a counter current f l ow DMPC setup.The error after the applicable corrections for the cup method was evaluated.The effects of relative humidity,temperature and pressure on the error of measured apparent and corrected permeances were also simulated.Favorable testing conditions to minimize the error are subsequently discussed.

2 Simulation

2.1 Assumptions

2D simulations for the experimental setup were developed employing the COMSOL Multiphysics 5.0 platform,an engineering modeling software based on the well-developed f i nite element method[Atluri(2005)].The consistency between experimental and simulation results for heat and mass transfer involving air f l ows using COMSOL has been reported in literatures[Bui,Chen,Nida,Chua and Ng(2015);Toujani,Djebali,Hassini,Azzouz and Belghith(2014);Lamloumi,Hassini,Lecomte-Nana,Elcafsi,Smith,Li,Huang,Ai and Tian(2014)].The following assumptions were made when developing these models:

1.Humid air approximates an ideal gas and the fluid f l ow of air is plug f l ow.

2.Water vapor transport is governed by the Fick’s law of diffusion and convection.The mass balance equation under isothermal steady state equilibrium is

wherecis the concentration of the water vapor(mol/m3),D is the water vapor diffusion coefficient(m2/s),uis the velocity vector(m/s).cis related to water vapor partial pressure and relative humidity as below:

Pwis the water vapor partial pressure(Pa),RH is the relative humidity andPsis the saturation pressure of water vapor(Pa).

The water vapor diffusion coefficient(D)is calculated from the empirical equation[Massman(1998)]:

wherePis the ambient pressure(Pa),Pois the standard atmospheric pressure(101325 Pa)andTis the temperature(K).

3.The water vapour permeance of the membrane is fixed and does not depend on the air RH.The water vapor transmission rate is proportional to the partial pressure difference between the two sides of the membrane,as in the below equation:wherekis the water vapor permeance of the membrane(mol/m2.s.Pa),ΔPwis transmembrane pressure andP1wandP2ware the water vapor partial pressure at two sides the membrane(Pa).

4.For the cup method,the changes in temperature and air gap due to the water evaporation/absorption are ignored.

Figure 2:Schematic diagrams for simulations of(a)cup setup and(b)DMPC setup

2.2 Cup method setup

The upright cup setup shown in Fig.1 a is modeled and simulated.The working membrane is modeled as a 10 mm long thin permeable barrier having constant water vapor permeance(k),placed between an outside layer of conditioned air and a still air layer in the cup as shown in Fig.2 a.In order to reduce the resistance to mass transfer,the outside air layer is moving atum/s.Water vapor transport in the outside air is governed by the Fick’s law of diffusion and convection.Its mass transfer process is depicted via equation(4).Water vapor transport in the still air gap inside the cup is governed only by the Fick’s law of diffusion.Its mass transfer process is simulated using equation(4)without the first term on the left hand side,which accounts for convective transport due to the velocity u.

The apparent water vapor permeance can be determined:

wherekappis the apparent water vapor permeance(mol/m2.s.Pa),PoutwandPinware the water vapor partial pressure of outside conditioned air and at the water/desiccant/salt solution surface respectively(Pa).

The error ofkappcompared withkis determined as:

The influence of the thickness of the still air gap(d),the conditioned air’sRH,temperature and pressure on theerroris studied judiciously.

2.3 DMPC method setup

The counter fl ow DMPC setup shown in Fig.1 c is simulated.The working membrane is a 10 mm long thin permeable barrier between a feed and a sweep flows,as shown in Fig.2 b.Flow velocities of feed air and sweep gas areufandusrespectively.Water vapor transport in both feed and sweep flows are governed by the Fick’s law of diffusion and convection and the mass balance equation for both flows is equation(4).

The apparent water vapor permeance can be determined:

在传统的生猪规模化养殖中,对于大量粪尿无法进行有效处理,通过对异位发酵床技术这一新型猪粪尿排泄物处理技术的应用,能在源头上对传统规模化猪场的粪尿污染问题进行处理,可以为养殖人员带来良好的经济效益。

wherePfeedinwandPfeedoutware the water vapor partial pressures at the inlet and outlet of the feed f l ow respectively(Pa),andPsweepinwandPsweepoutware the water vapor partial pressures at the inlet and outlet of the sweep f l ow respectively(Pa).The error ofkappcompared withkis determined using equation(9).The inf l uence of the sweep gas velocity,pressure,conditioned air’sRHand temperature on the error is analyzed and quantified.

3 Results and discussion

3.1 Transmembrane water vapor partial pressure,

Figure 3:(a)water vapor partial pressure along the y axis and(b)transmembrane pressure in dry cup method setup.

Fig.3 a shows the water vapor partial pressure along theyaxis in a dry cup method setup at different membrane permeances.The relative humidity of the outside conditioned air is set at 70%.A desiccant like silica gel or calcium chloride is modelled in this setup and the relative humidity at the desiccant surface is set at 0%.The temperature is 298K.The outside air layer is moving atu=10 m/s.The water vapor permeance of the membrane(k)is varied from 10−8to 5 10−6mol/m2.s.Pa.The result shows that there is an accumulation of water vapor on the inner side of the membrane,causing an increase in water vapor partial pressure near the membrane.This accumulation is small with low water vapor permeance spanning from 10−8to 10−7mol/m2.s.Pa and increases sharply with higher water vapor permeance.There is still a significant amount of water vapor dissipation on the outer side of the membrane when water vapor permeance is high,causing a drop in pressure near the membrane,even though the conditioned air is moving at 10 m/s,4 times higher recommended velocity in ASTM 96[ASTM(2014)].As a result,the actual water vapor transmembrane pressure(ΔPw)is much lower than the water vapor pressure difference between the desiccant surface and outside conditioned air.

Fig.3 b shows the dependence of the water vapor transmembrane pressure on the membrane permeance and the thickness of the still air(d)between the desiccant surface and the membrane.The actual water vapor transmembrane pressure decreases with higher water vapor permeance and thicker air gap.

Fig.4 a shows the water vapor partial pressure along theyaxis in a DMPC setup.The relative humidity of the feed air and the sweep gas is set at 70%and 0%,respectively.The temperature is 298K.The feed air and the sweep gas are moving at 10 and 0.5 m/s,respectively.The water vapor permeance of the membrane(k)varies from 10−8to 5 10−6mol/m2.s.Pa.Similar as the cup method,there are significant water vapor accumulation and dissipation caused by the resistance of air phases on both sides of the membrane in this DMPC setup.This results in a lower water vapor transmembrane pressure compared with the mean water vapor partial pressure difference between the feed air and the sweep gas.As shown in Fig.4 b,the actual water vapor transmembrane pressure decreases with higher water vapor permeance and slower sweep gas velocity.

Figure 4:(a)water vapor partial pressure along the y axis and(b)transmembrane pressure in a dynamic moisture permeation cell setup.

3.2 Apparent and actual permeance

Figure 5:Comparision between apparent and actual permeances in(a)cup setup and(b)DMPC method

During the practical testing of both methods,water vapor permeance is determined based on the assumption that the mass transport resistance of the air on both sides of the membrane is negligible.This leads the apparent permeance computed using equations(8)and(10)to be lower than the actual water vapor permeance.The underestimation of water vapor permeance is shown in Fig.5.From the two graphs,it is seen that the apparent permeance diverges from the actual permeance as membrane permeance increases.Further,the divergence increases with higher air gap in the cup method and slower sweep velocity in the DMPC method.It is therefore apparent that the resistances to mass transfer in the gas boundary layers on both sides of the membrane cannot be neglected,especially in case of highly water permeable membranes.Simplifying calculations in practical tests by ignoring this resistance will cause error of the measurement.The errors determined by equation(9)for the two methods are shown in Fig.6.

Figure 6:Error caused by ignorance of mass transport resistance of air boundary layers in(a)the cup method and(b)the DMPC method.

As shown in Fig.6,both methods potentially evolve large errors if the resistances of the air boundary layers are not taken into account.The error increases with higher water vapor permeance.A thicker air gap in the cup method and a lower sweep velocity in the DMPC method also cause higher error.

3.3 Correction for resistance of still air

According to ASTM E96-14[ASTM(2014)],all measurements that result in permeance values of more than 2-perms(6.33×10−9mol/m2s Pa)require corrections.The correction for resistance due to the still air is based on the water vapor permeance of air(kair,mol/m2.s.Pa):

The water vapor permeance with the correction(kcor,mol/m2.s.Pa)is determined as follows:

For the inverted cup method in which there is no still air gap,the measured water vapour permeance would be similar to the corrected water vapour permeance values shown in Fig.7 a.For the convenience of not applying the corrections,the inverted cup method can be used.

Figure 8:RH change between sweep gas inlet and outlet

In the DMPC method,there has been no reported literature on such a kind of correction for the DMPC method.Therefore,only by increasing the sweeping velocity,the error can be reduced as shown in Fig.6 b.However,the change in RH also becomes smaller as shown in Fig.8.The smaller change in water content can magnify the instrument error.This means that there is always a tradeoff between the error caused by the resistance of the air boundary layers and the instrument error with the increase of sweeping velocity.Highly accurate and precise instrument is needed for the DMPC method.However,this does not ensure that the measurement error of the DMPC method is lower than that of the cup method.This is because with the sweep velocity of 10 m/s,at which the sweep stream’s RH is almost unchanged(as shown in Fig.8),the error of the DMPC method(shown in Fig.6 b)is still higher than the error after correction of the cup method with fresh velocity higher 10 m/s(shown in Fig.7 c.).

3.4 Effects of RH,temperature and pressure on measurement error

Adopting the assumption that the water vapor permeance of a membrane does not change with RH of the air,the simulation results show that the measurement error also does not depend on the difference in RH values between both sides of the membranes.As shown in Fig.9 a,the curves for the error ofkappandkcoroverlap at differing RH values outside the cup while the RH inside the cup controlled using desiccant is kept at 0%.The result is also applicable for the wet-cup method.

The independence of measurement error from RH is also observed in the case of the DMPC method as shown in Fig.9 b.The error curves overlap at different feed RH values while the sweep inlet RH is kept at 0%.However,the lowering of the feed RH leads to small RH changes in the sweep stream as shown in Fig.9 c,causing higher instrument errors.Therefore,when the DMPC method is to be considered,a higher RH difference between the two f l ows is desirable.

Figure 9:Effects of RH(a,b and c),temperature(d,e)and pressure(f,g)on measurement errors in the cup(a,d,f)and DMPC(b,e and g)methods and sweep gas’s RH change in DMPC method(c).

Adopting the assumption that the water vapor permeance of a membrane is independent of temperature,it follows that a change in temperature just affects the diffusion of water vapor in the air.In both methods,increasing process temperature leads to higher water vapor diffusion and a lower resistance of the boundary air layer.Consequently,lower errors are achieved,as shown in Fig.9 d and e.However,when the temperature increases from 293 K to 323 K,the error ofkappfor both the cup and DMPC methods decrease only slightly while the error ofkcoris almost unchanged for the cup method.

Pressure affects the water vapor diffusion in air as apparent through equation(6).A decrease in pressure will result in higher diffusion effect while lowering the resistance of air to water vapor transport.In both methods,higher errors are obtained under high pressure conditions as shown in Fig.9 f and g.This result is consistent with the report that concentration polarization effects cannot be neglected in high pressure applications[Lüdtke,Behling and Ohlrogge(1998)].

4 Conclusions

In this work,two commonly used techniques,the cup and DMPC methods,were studied by numerical analysis to evaluate water vapor permeance of highly permeable membranes.Under room condition and without proper corrective intervention,both methods give high errors due to the effect of water vapor transport resistance of the boundary air layers on both sides of the membrane not being considered.Because the resistance of still air can be determined,the cup method provides a corrected water vapor permeance value closer to the actual one without compro-mising instrument error.Highly accurate and precise instrument does not ensure a lower error in the DMPC method than the cup method.Both methods can be markedly improved by lowering process pressure.

Acknowledgement: The authors gratefully acknowledge the financial support provided by both Agency for Science,Technology and Research(A*Star)and Ministry of National Development(MND)through their Green Building Joint Grant(no:112 176 0023)and National Research Foundation(NRF)Competitive Research Programme(CRP)NRF2011NRF_CRP003_003.

ASTM(2009): ASTM F2298-03(2009)e1:Standard Test Methods for Water Vapor Diffusion Resistance and Air Flow Resistance of Clothing Materials Using the Dynamic Moisture Permeation Cell(Withdrawn 2015).West Conshohocken,Penn.ASTM International.

ASTM(2014): ASTM E96/E96M-14:Standard Test Methods for Water Vapor Transmission of Materials.West Conshohocken,Penn.ASTM International.

Atluri,S.N.(2005):Methods of Computer Modeling in Engineering&the Sciences,Volume 1.Tech Science Press.

Bui,T.D.;Chen,F.;Nida,A.;Chua,K.J.;Ng,K.C.(2015):Experimental and modeling analysis of membrane-based air dehumidification.Sep.Purif.Technol.,vol.144,pp.114-122.

Gennadios,A.;Weller,C.L.;Gooding,C.H.(1994):Measurement errors in water vapor permeability of highly permeable,hydrophilic edible f i lms.J.Food Eng.,vol.21,pp.395-409.

Hu,Y.;Topolkaraev,V.;Hiltner,A.;Baer,E.(2001):Measurement of water vapor transmission rate in highly permeable films.J.Appl.Polym.Sci.,vol.81,pp.1624-1633.

Huang,J.(2008):A Device for Characterizing Water Vapor Permeability of Polymer Membranes.Int.J.Polym.Anal.Charact.,vol.13,pp.37-48.

Huang,J.;Qian,X.(2008):Comparison of Test Methods for Measuring Water Vapor Permeability of Fabrics.Text.Res.J.,vol.78,pp.342-352.

Krull,F.;Fritzmann,C.;Melin,T.(2008):Liquid membranes for gas/vapor separations.J.Membr.Sci.,vol.325,pp.509-519.

Lamloumi,R.;Hassini,L.;Lecomte-Nana,G.;Elcafsi,M.;Smith,D.;Li,W.;Huang,H.;Ai,B.;Tian,Y.(2014): Modeling of Hydro-Viscoelastic State of Deformable and Saturated Product During Convective Drying.CMC:Computers,Materials&Continua,vol 23,pp.137-151.

Lin,H.;Thompson,S.M.;Serbanescu-Martin,A.;Wijmans,J.G.;Amo,K.D.;Lokhandwala,K.A.;Merkel,T.C.(2012):Dehydration of natural gas using membranes.Part I:Composite membranes.J.Membr.Sci.,vol 413–414,pp.70-81.

Lüdtke,O.;Behling,R.-D.;Ohlrogge,K.(1998):Concentration polarization in gas permeation.J.Membr.Sci.,vol.146,pp.145-157.

Massman,W.(1998):A review of the molecular diffusivities of H2O,CO2,CH4,CO,O3,SO2,NH3,N2O,NO,andNO2inair,O2andN2nearSTP.Atmos.Environ.,vol.32,pp.1111-1127.

McCullough,E.A.;Kwon,M.;Shim,H.(2003):A comparison of standard methods for measuring water vapour permeability of fabrics.Meas.Sci.Technol.,vol.14,pp.1402-1408.

McHugh,T.H.;Avena-Bustillos,R.;Krochta,J.(1993): Hydrophilic Edible Films:Modif i ed Procedure for Water Vapor Permeability and Explanation of Thickness Effects.J.Food Sci.,vol.58,pp.899-903.

Metz,S.;Van De Ven,W.;Mulder,M.;Wessling,M.(2005):Mixed gas water vapor/N2 transport in poly(ethylene oxide)poly(butylene terephthalate)block copolymers.J.Membr.Sci.,vol.266,pp.51-61.

Metz,S.;Van de Ven,W.;Potreck,J.;Mulder,M.;Wessling,M.(2005):Transport of water vapor and inert gas mixtures through highly selective and highly permeable polymer membranes.J.Membr.Sci.,vol.251,pp.29-41.

Roy,S.;Hussain,C.M.;Mitra,S.(2013):Poly(acrylamide-co-acrylic acid)hydrophilization of porous polypropylene membrane for dehumidif i cation.Sep.Purif.Technol.,vol.107,pp.54-60.

Scovazzo,P.(2010): Testing and evaluation of room temperature ionic liquid(RTIL)membranes for gas dehumidif i cation.J.Membr.Sci.,vol.355,pp.7-17.

Sijbesma,H.;Nymeijer,K.;van Marwijk,R.;Heijboer,R.;Potreck,J.;Wessling,M.(2008):Flue gas dehydration using polymer membranes.J.Membr.Sci.,vol.313,pp.263-276.

Toujani,M.;Djebali,R.;Hassini,L.;Azzouz,S.;Belghith,A.(2014):Hydrothermo-viscoelastic Based Finite Element Modeling of Apple Convective Drying Process.CMES:Computer Modeling in Engineering&Sciences,vol.98,pp.469-485.

Xing,R.;Rao,Y.;TeGrotenhuis,W.;Canfield,N.;Zheng,F.;Winiarski,D.W.;Liu,W.(2013): Advanced thin zeolite/metal f l at sheet membrane for energy efficient air dehumidif i cation and conditioning.Chem.Eng.Sci.,vol.104,pp.596-609.

Yang,B.;Yuan,W.;Gao,F.;Guo,B.(2015):A review of membrane-based air dehumidification.Indoor Built Environ.,vol.24,no.1,pp.11-26.

Zhang,L.-Z.(2006):Evaluation of moisture diffusivity in hydrophilic polymer membranes:A new approach.J.Membr.Sci.,vol.269,pp.75-83.

Zhang,L.;Jiang,Y.(1999):Heat and mass transfer in a membrane-based energy recovery ventilator.J.Membr.Sci.,vol.163,pp.29-38.

1Corresponding author E-mail:mpebuid@nus.edu.sg/buiducthuan@gmail.com

2National University of Singapore,Department of Mechanical Engineering,9 Engineering Drive 1,Singapore 117575,Singapore

猜你喜欢
粪尿排泄物猪粪
陕西省畜禽粪尿养分资源及耕地负荷分析
猪粪配施化肥对侵蚀林地土壤团聚体及其有机碳分布的影响
猪粪变有机肥一年卖了3个亿
猪粪中添加腐殖酸添加剂可降低粪便中的臭气
猫咪为何埋便便
异位发酵床猪粪尿处理技术综述
延边地区畜禽粪便污染现状及产沼气潜力分析
菌渣猪粪还田下麦稻重金属富集特征及风险评价
丹麦惊现700年前的厕所
高床养羊的优势